Vol. 2, Article 2 Neurographics logo Erickson

 

Motion

To this point, we have focused on the images, and not the patient. Because patients are not always able to cooperate, motion can occur. Since the computation is based on tracking a certain sample of tissue through time, patient motion can result in errors in perfusion images. On the next screen is a movie loop of a patient that moved during the perfusion sequence.

Motion During Perfusion Sequence

Correction for Motion

There are computer techniques that can align volumetric images with each other. This can be applied to perfusion images to correct for motion. In the following slide, a data set with no perceptible motion was taken, and each timepoint after timepoint 10, was rotated 1 degree (total rotation = 30 degrees at timepoint 40). An internally developed application was then applied that retrospectively aligns images.

CBV Map From 1 Degree Potation Per Image

No Motion Correction With Motion Correction Original
Figure 10 Figure 11 Figure 12

Motion Correction

While retrospective motion correction produces notable improvement for marked motion, the results are more variable for minimal motion. The following shows the results when each image is rotated just 0.1 degree, for a total of 3 degrees rotation on the final image. In this case, there improvement, but in other cases, the registration process did not improve the CBV map.

CBV Map From 0.1 Degrees Rotation Per Image

No Motion Correction With Motion Correction Original
Figure 13 Figure 14 Figure 15

 



Copyright © 2002. All rights reserved.
Use of this material without written consent of the author is prohibited.

Return to Neurographics Main Page

 



Copyright © 2002. All rights reserved.
Use of this material without written consent of the author is prohibited.

Return to Neurographics Main Page